For the temperature calibration professional that wants a highly accurate, easy-to-use, single function RTD temperature calibrator the Fluke 712B is the ideal test tool.
Fluke 712B Comes Supplied With:
- Magnetic hanging tool
- Batteries
- Manual
- Traceable calibration certificate
- Test leads
General Specifications |
Maximum voltage applied between any terminal and earth ground or between any two terminals |
30 V |
Operating temperature |
-10 °C to 50 °C |
Storage temperature |
- 30 °C to 60 °C |
Operating altitude |
2,000 meters |
Storage altitude |
12,000 meters |
Relative humidity (% RH operating without condensation) |
Non condensing |
90% (10 °C to 30 °C) |
75% (30 °C to 40 °C) |
45% (40 °C to 50 °C) |
(Without condensation) |
Vibration requirements |
MIL-T-28800E, Class 2 |
Drop text requirements |
1 meter |
IP Rating |
IEC 60529: IP52 |
Electromagnetic environment |
IEC 61326-1, Portable |
Safety |
IEC 61010-1, Max 30 V to earth, Pollution Degree 2 |
Power supply |
4 AA NEDA 1.5A IEC LR6 batteries |
Size (H x W x L) |
52.5 mm x 84 mm x 188.5 mm |
Weight |
515 g |
DC mA Measurement |
Resolution |
0-24 mA |
Range |
0.001 mA |
Accuracy (% of reading + counts) |
0.010% + 2 μA |
Temperature coefficient |
± (0.002% of reading + 0.002% of range) /°C ( <18 °C or >28 °C) |
Ohms Measurement |
Ohms range |
Accuracy (% of reading +counts) |
0.00 Ω to 400.00 Ω |
0.015% + 0.05 Ω |
400.0 Ω to 4000.0 Ω |
0.015% + 0.5 Ω |
Note: Read accuracy is based on 4-wire input. For 3-wire ohm measurements, assuming all three leads are matched, add 0.05 Ω (0.00 Ω~400.00 Ω), 0.2 Ω (400.0 Ω~4000.0 Ω) to the specifications. |
Temperature coefficient |
± (0.002% of reading + 0.002% of range) /°C ( <18 °C or >28 °C) |
Ohms Source |
Ohms range |
1.0 Ω to 400.0 Ω |
1.00 Ω to 400.00 Ω |
400.0 Ω to 1500.0 Ω |
1500.0 Ω to 4000.0 Ω |
Excitation current from measurement device |
0.1 mA to 0.5 mA |
0.5 mA to 3 mA |
0.05 mA to 0.8 mA |
0.05 mA to 0.4 mA |
Accuracy (% of reading + counts) |
0.015% + 0.1 Ω |
0.015% + 0.05 Ω |
0.015% + 0.5 Ω |
0.015% + 0.5 Ω |
Resolution |
0.00 Ω to 400.00 Ω |
0.01 Ω |
400.0 Ω to 4000.0 Ω |
0.1 Ω |
Temperature coefficient |
± (0.002% of reading + 0.002% of range) /°C ( <18 °C or >28 °C) Supports pulsed transmitters and PLCs with pulse times as short as 5 ms |
RTD Input and Output |
RTD Type (α) |
Range (°C) |
Measure (°C) |
Source (°C) |
1-year |
2-year |
Source current |
1-year |
2-year |
10Ω Pt(385) |
-200 to 100 °C |
1.5 °C |
3 °C |
1 mA |
1.5 °C |
3 °C |
100 to 800 °C |
1.8 °C |
3.6 °C |
1 mA |
1.8 °C |
3.6 °C |
50Ω Pt(385) |
-200 to 100 °C |
0.4 °C |
0.7 °C |
1 mA |
0.4 °C |
0.7 °C |
100 to 800 °C |
0.5 °C |
0.8 °C |
1 mA |
0.5 °C |
0.8 °C |
100 Ω Pt(385) |
-200 to 100 °C |
0.2 °C |
0.4 °C |
1 mA |
0.2 °C |
0.4 °C |
100 to 800 °C |
0.015% +0.18 °C |
0.03% +0.36 °C |
0.015% +0.18 °C |
0.03% +0.36 °C |
200 Ω Pt(385) |
-200 to 100 °C |
0.2 °C |
0.4 °C |
500 μA |
0.2 °C |
0.4 °C |
100 to 630 °C |
0.015% +0.18 °C |
0.03% +0.36 °C |
0.015% +0.18 °C |
0.03% +0.36 °C |
500 Ω Pt(385) |
-200 to 100 °C |
0.3 °C |
0.6 °C |
250 μA |
0.3 °C |
0.6 °C |
100 to 630 °C |
0.015% +0.28 °C |
0.03% +0.56 °C |
0.015% +0.28 °C |
0.03% +0.56 °C |
1000 Ω Pt(385) |
-200 to 100 °C |
0.2 °C |
0.4 °C |
250 μA |
0.2 °C |
0.4 °C |
100 to 630 °C |
0.015% +0.18 °C |
0.03% +0.36 °C |
0.015% +0.18 °C |
0.03% +0.36 °C |
100 Ω Pt(3916) |
-200 to 100 °C |
0.2 °C |
0.4 °C |
1 mA |
0.2 °C |
0.4 °C |
100 to 630 °C |
0.015% +0.18 °C |
0.03% +0.36 °C |
0.015% +0.18 °C |
0.03% +0.36 °C |
100 Ω Pt(3926) |
-200 to 100 °C |
0.2 °C |
0.4 °C |
1 mA |
0.2 °C |
0.4 °C |
100 to 630 °C |
0.015 % +0.18 °C |
0.03% +0.36 °C |
0.015% +0.18 °C |
0.03% +0.36 °C |
10 Ω Cu(427) |
-100 to 260 °C |
1.5 °C |
3 °C |
1 mA |
1.5 °C |
3 °C |
120 Ω Ni(672) |
-80 to 260 °C |
0.15 °C |
0.3 °C |
1 mA |
0.15 °C |
0.3 °C |
50 Ω Cu(427) |
-180 to 200 °C |
0.4 °C |
0.7 °C |
1 mA |
0.4 °C |
0.7 °C |
100 Ω Cu(427) |
-180 to 200 °C |
0.2 °C |
0.4 °C |
1 mA |
0.2 °C |
0.4 °C |
YSI400 |
15 to 50 °C |
0.2 °C |
0.4 °C |
250 μA |
0.2 °C |
0.4 °C |
1. Sensor inaccuracies not included. 2. Resolution: 0.1 °C. 3. Read accuracy is based on 4-wire input. For 3-wire RTD measurements, assuming all three RTD leads are matched, add 1.0 °C (Pt10 and Cu10), 0.6 °C (Pt50 and Cu50), 0.4 °C (Other RTD types) to the specifications. 4. Source Accuracy in source mode is based on 0.5 mA~3 mA (1.00 Ω~400.00 Ω), 0.05 mA~0.8 mA (400.0 Ω~1500.0 Ω), 0.05 mA~0.4 mA (1500.0 Ω ~4000.0 Ω), excitation current (0.25 mA for Pt1000 range). 5. Temperature Coefficient: ±0.05 °C /°C for measure, ± 0.05 °C /°C (<18 °C or >28 °C) for source. 6. Supports pulsed transmitters and PLCs with pulse times as short as 5 ms. |
|